8 Troubleshooting


Symptom

Possible Cause

Recommended Action

Reading does not

reflect expected values

Sensor was not calibrated at the

pressure, flow rate and temperature

anticipated in the sample gas stream

Recalibrate the analyzer

Oxygen reading drifts

toward zero or

significant number of

turns of the span

control adjustment is

required to calibrate the

analyzer.

Indication sensor is nearing the end of

its useful life

Replace sensor, see Section 6 -

Maintenance.


Slow response time

Erratic oxygen reading

No oxygen reading


Liquid covering sensing membrane

Presence of interference gases.

Unauthorized maintenance

Defective electrical connection

Sensor failure


Gently remove with alcohol and

lint free towel.
Consult factory, replace sensor,

see Section 6 - Maintenance.




Use voltmeter and determine uA or

mV output and contact factory.


High oxygen reading


Inadequate control of pressure and

flowrate

Abnormality in span gas


See Section 5 - Operation,
Getting Started,
Control of Pressure and Flow

Qualify source