Advanced instrtuments inc, 5 operation - appendix a – Analytical Industries GPR-3100 Series Oxygen Purity Analyzer User Manual
Page 40

Advanced Instrtuments Inc.
5 Operation - Appendix A
* MAIN MENU Sample
Sample
Span
Zero
Alarm
System
Auto Range 90 to 100 %
75F 101 Kpa 02/10/2004 17:32:48
99.55 %
MAIN MENU Sample
SPAN Sample
* MAIN MENU Sample
Sample
* Factory Default (1)
Sample
(1) SPAN menu, Factory Default option -
* Span
Calibrate
Span
Purpose is to reset the microprocessor as the first step in
Zero
Zero
troubleshooting perceived abnormalities. The span value
Alarm
Alarm
(which may be compromised) is reset using an average
System
System
sensor output value to temporarily assess the analyzer's
performance. Calibration is required upon completion.
Auto Range 90 to 100 %
Auto Range 90 to 100 %
Auto Range 90 to 100 %
75F 101 Kpa 02/10/2004 17:32:48
75F 101 Kpa 02/10/2004 17:32:48
75F 101 Kpa 02/10/2004 17:32:48
Move cursor (*) to Span using DOWN ARROW key
Move cursor (*) to Factory Default using DOWN ARROW key
Press ENTER to select
Press ENTER to select
Activates (1) and returns to MAIN MENU
Note: Returns to MAIN MENU if no entry within 30 seconds
SPAN Sample
Span Gas Sample
100.00 % Sample
* MAIN MENU Sample
Factory Default
Enter as PPB
*
Sample
* Calibrate
Enter as PPM
Press Up or Down
Span
* Enter as %
to change value (2)
Zero
Enter to Save
Alarm
ESC to Return
System
Auto Range 90 to 100 %
Auto Range 90 to 100 %
Auto Range 90 to 100 %
Auto Range 90 to 100 %
75F 101 Kpa 02/10/2004 17:32:48
75F 101 Kpa 02/10/2004 17:32:48
75F 101 Kpa 02/10/2004 17:32:48
75F 101 Kpa 02/10/2004 17:32:48
Move cursor (*) to Calibrate using DOWN ARROW key
Move cursor (*) to desired Man Range using DOWN ARROW key
(2) To change value of 1st digit.
Press ENTER to select
Press ENTER to select
Press ENTER to advance cursor (*) to 2nd digit
Returns to MAIN MENU
Press Up or Down to change value
Repeat as necessary for remaining digits
99.55 %
99.55 %
99.55 %
5_ops_3100_menu 11/22/2005 Page 2 of 7