Acquiring spectra, Proc – Ocean Optics MMS Raman User Manual

Page 12

background image

3: Operating the Spectrometer

Acquiring Spectra

Procedure

1. Remove the sample holder cover.

2. Place the sample cuvette in the sample holder and adjust its distance from the lens using the Z-

axis knob. Then, replace the cover.

3. Turn the laser safety key to the On position (vertical) to activate the laser. The laser is not active

at this point. A signal is sent to the spectrometer’s processor requesting activation of the laser.
The LED on the spectrometer lights. Two seconds after the processor receives the laser activation
request, the processor sends the activation signal to the laser controller. The laser is now
operating.

Note

If you attempt to take a scan with the laser off, a warning appears in SpectraSuite:

4. If you have not already done so, start the SpectraSuite software application.

SpectraSuite should now recognize the spectrometer (a picture of the spectrometer appears in the
Data Sources pane of the SpectraSuite window).