System compatibility, Pin accessory connector pinout – Ocean Optics QE65000 User Manual
Page 31
B: Specifications
220-00000-000-02-201303
25
Specification
Value
Gratings
14 gratings available
Entrance aperture
5, 10, 25, 50, 100 or 200 µm wide slits
Order-sorting filters
6 OF series available
Focal length (input)
f/4, 101 mm
Optical resolution
(FWHM)
Depends on grating and size of entrance aperture
Stray light
<0.08% at 600 nm; 0.4% at 435 nm
Dynamic range
7.5 x 10
9
(over integration range); 25000:1 for a single integration
period
Fiber optic connector
SMA 905 to single-strand optical fiber (0.22 NA)
Integration time
8 ms to 15 minutes
Interfaces
USB 2.0, 480 Mbps (USB 1.1 compatible); RS-232 (3-wire)
System Compatibility
You can use the QE65000’s USB connectivity with any computer that meets the requirements for the
spectrometer operating software being used (Windows 2000/XP, Mac OS X and Linux). See
Alternately, the QE65000 has a serial port for connecting to PCs, PLCs, and other devices with a
Windows 32-bit operating system that support the RS-232 communication protocol. However, this
connection method requires an external power supply to power the QE65000, the HR4000 Breakout Box,
and a serial cable.
30-Pin Accessory Connector Pinout
The QE65000 features a 30-pin Accessory Connector, located on the side of the unit as shown: