B: Specifications

220-00000-000-02-201303

25

Specification

Value

Gratings

14 gratings available

Entrance aperture

5, 10, 25, 50, 100 or 200 µm wide slits

Order-sorting filters

6 OF series available

Focal length (input)

f/4, 101 mm

Optical resolution
(FWHM)

Depends on grating and size of entrance aperture

Stray light

<0.08% at 600 nm; 0.4% at 435 nm

Dynamic range

7.5 x 10

9

(over integration range); 25000:1 for a single integration

period

Fiber optic connector

SMA 905 to single-strand optical fiber (0.22 NA)

Integration time

8 ms to 15 minutes

Interfaces

USB 2.0, 480 Mbps (USB 1.1 compatible); RS-232 (3-wire)

System Compatibility

You can use the QE65000’s USB connectivity with any computer that meets the requirements for the
spectrometer operating software being used (Windows 2000/XP, Mac OS X and Linux). See

About Ocean

Optics Software

.

Alternately, the QE65000 has a serial port for connecting to PCs, PLCs, and other devices with a
Windows 32-bit operating system that support the RS-232 communication protocol. However, this
connection method requires an external power supply to power the QE65000, the HR4000 Breakout Box,
and a serial cable.

30-Pin Accessory Connector Pinout

The QE65000 features a 30-pin Accessory Connector, located on the side of the unit as shown: