2: Installing the OTS

Incorrectly Adjusted Integration Time: Integration time set too low

5. When finished with setting parameters, close the Scan Parameters window by clicking the red

in the upper right corner of the Scan Parameters window (not the OTS window!).

6. Calibrate OTS. To do this, click the drop-down menu in the Reference box (in the upper right of

the Main screen) to choose to calibrate to either Fused SiO2 or Air. Then, click the

button. The following popup dialog box appears:

7. If you are calibrating using fused SiO2 as the standard, place the thin SiO2 sample on the stage.

If you are calibrating using air as the standard, place no sample on the stage.

Click OK, and leave the calibration piece in place until the following dialog box appears:

8. Click OK to return to the Main screen.