Algorithms used, Ph calculation, Temperature compensation – Ocean Optics Smart pH Cuvettes User Manual

Page 7: Resetting pk and slope

background image

Smart Cuvettes Instructions

SC-PH-VIS1M-SAM-01-1109

7

Algorithms Used

pH Calculation



Sample

pH

Sample

Abs

Abs

Abs

Slope

pK

pH

11

log

*

…where Abs

Sample

is the sample absorbance at 620nm with baseline correction, and Abs

pH11

is the

absorbance at pH 11 at 620nm with baseline correction.

Temperature Compensation

When you select Use Factory Calibration in SpectraSuite, the value for pK is adjusted via the van’t
Hoff equation based on the current temperature you entered:









1

2

1

2

1

1

*

480

1

2

1

1

*

480

1

2

log

log

T

T

T

T

e

pH

pH

e

pK

pK

Resetting pK and Slope

An x-y plot is made using data obtained from intermediate buffers 5 through 8. The x-axis is of the
term:



Sample

pH

Sample

Abs

Abs

Abs

11

log

…for each of the buffers. The y-axis shows the pH value of the buffers. This generates a plot such as
the one shown below: