beautypg.com

Ocean Optics Smart pH Cuvettes User Manual

Page 6

background image

Smart Cuvettes Instructions

6

SC-PH-VIS1M-SAM-01-1109

2. Press the Run/Stop button to toggle data acquisition appearing in the lower table on the

screen. Data is recorded at the time interval you specify in the Time Interval (sec) field.

3. Click the Reset button to clear the table and restart the run time.

4. Click the Export button to open a window to save your data in a format that can be opened

with Microsoft Excel or a text program such as Wordpad. The exported data file contains all
of the variables that you have entered and have been calculated, along with a time stamp for
data acquisition and save, the time-resolved pH data, and complete spectra for all reference
and calibration buffers used.

5. Click the Export Calibration button to open a window to save your calibration data. This will

create a file containing the reference spectra and other variables that can later be loaded via
the Calibration Wizard, allowing for very quick setup.