NeoFox Communication Interfaces

30 013-20000-010-04-201211

Single Point Temperature

float

X

≤ 200

188


Temperature (in degrees C) to use for next SPR calculation. Most users will want to set this to the current value of
Temperature read by the instrument while in the SPR environment.

Single Point Tau

float

X ≤ 10.0

186


Tau (in microseconds) to use for next SPR calculation. Most users will want to set this to the current value of Tau
read by the instrument while in the SPR environment.

Single Point Oxygen

float

0 ≤ X

187


Percent Oxygen (in percent of 1 ATM) to use for next SPR calculation. Most users will want to set this to the known
value of percent oxygen in the SPR environment. In ambient air, at sea level, this will be equal to 20.9.

Single Point Calculate

N/A

N/A

189


Writing any value to this parameter will result in a single point reset with the single point reset parameter values
listed above. The new single point coefficients will be calculated and the device will be put into single point mode
(as opposed to multipoint mode). The new value will not necessarily be set in flash memory until the user manually
issues a flash write command.