NeoFox Communication Interfaces
013-20000-010-04-201211 15
0-5V Data Source
Char
Enum
212
468
NEOFOX_AOUT_VOLTAGE_SOURCE
DeviceSetParameter
DeviceGetParameter
The 0-5V Data Source parameter is one of the 0-5V analog output settings. The value of this particular parameter sets the
source value for the following formula, which determines the 0-5V analog output voltage. In the formula, the source value
and bound parameters are considered dimensionless scalars – that is, only the actual values matter. The units are irrelevant.
More information about this parameter and how to set up the analog outputs is available in the user manual.
Output (Volts) = ( Source Value – Lower Bound Value ) / (Upper Bound Value – Lower Bound Value) * 5 Volts
Value
Description
0
Output is disabled ( 0V )
1
Percent Oxygen (in % of 1 ATM)
2
Sensor Temperature (in degrees C)
3
Ambient Pressure (in kPa)
4
Tau (in microseconds)
5
Converted O2 (in the Converted O2 units selected)
6
7
Analog Value 1
Analog Value 2
4-20mA Data Source
Char
Enum
213
469
NEOFOX_AOUT_CURRENT_SOURCE
DeviceSetParameter
DeviceGetParameter
The 4-20mA Data Source parameter is one of the 4-20mA analog output settings. The value of this particular parameter sets
the source value for the following formula, which determines the 4-20mA analog output current. In the formula, the source
value and bound parameters are considered dimensionless scalars – that is, only the actual values matter. The units are
irrelevant. More information about this parameter and how to set up the analog outputs is available in the user manual.
Output (mA) =[16mA*(Source Value – Lower Bound Value) / (Upper Bound Value – Lower Bound Value)]+4mA
The 4-20mA Data Source parameter is one of the 4-20mA analog output settings. The value of this particular parameter sets
the source value for the following formula, which determines the 4-20mA analog output current. In the formula, the source
value and bound parameters are considered dimensionless scalars – that is, only the actual values matter. The units are
irrelevant. More information about this parameter and how to set up the analog outputs is available in the user manual.
Output (mA) =[16mA*(Source Value – Lower Bound Value) / (Upper Bound Value – Lower Bound Value)]+4mA
Value
Description
0
Output is disabled ( 4 mA )
1
Percent Oxygen (in % of 1 ATM)
2
Sensor Temperature (in degrees C)
3
Ambient Pressure (in kPa)
4
Tau (in microseconds)
5
Converted O2 (in the Converted O2 units selected)
6
7
Analog Value 1
Analog Value 2
Salinity Correction Factor
Float
0 ≤ X
218
492
NEOFOX_SALINITY_CORRECTION
DeviceSetParameterFloat
DeviceGetParameterFloat
The Salinity Correction Factor is used in the conversion partial pressure to dissolved oxygen. More information about how
this parameter is used is available in the user manual.