Configuring calibration with the dh-2000-cal, Configuring the dh-2000-cal – Ocean Optics DH-2000-CAL User Manual
Page 20
3: Operating Instructions
12
000-10000-030-02-1209
Configuring Calibration with the DH-2000-CAL
The following sections provide instructions on configuring calibration with the DH-2000-CAL.
Configuring the DH-2000-CAL
►
Procedure
Follow the steps below to configure the DH-2000-CAL:
1. If using a bare fiber with the DH-2000-CAL:
a.
Screw the fiber end onto the SMA barrel adapter until secure.
b.
Insert the SMA barrel adapter completely into the optical input on the DH-2000-CAL.
If using a CC-3 cosine corrector with the DH-2000-CAL:
a.
Screw the CC-3 cosine corrector onto one end of the fiber until the connection is tight.
b.
Insert the CC-3 cosine corrector completely into the input on the DH-2000-CAL.
2. Use the Allen wrench to tighten the setscrew on the optical input of the DH-2000-CAL.
3. Screw the other end of the fiber into the SMA 905 Connector of the spectrometer.
4. Plug the black power cord into the back of the DH-2000-CAL.
5. Plug the other end of the power cord into a standard outlet.
6. Turn the lamp power on using the black switch on the rear of the DH-2000-CAL.
7. Switch the individual lamps on using the buttons on the front of the DH-2000-CAL.
8. Allow the lamp to warm up for at least 20 minutes before proceeding.