3: Operating Instructions

12

000-10000-030-02-1209

Configuring Calibration with the DH-2000-CAL

The following sections provide instructions on configuring calibration with the DH-2000-CAL.

Configuring the DH-2000-CAL

Procedure

Follow the steps below to configure the DH-2000-CAL:

1. If using a bare fiber with the DH-2000-CAL:

a.

Screw the fiber end onto the SMA barrel adapter until secure.

b.

Insert the SMA barrel adapter completely into the optical input on the DH-2000-CAL.

If using a CC-3 cosine corrector with the DH-2000-CAL:

a.

Screw the CC-3 cosine corrector onto one end of the fiber until the connection is tight.

b.

Insert the CC-3 cosine corrector completely into the input on the DH-2000-CAL.

2. Use the Allen wrench to tighten the setscrew on the optical input of the DH-2000-CAL.

3. Screw the other end of the fiber into the SMA 905 Connector of the spectrometer.

4. Plug the black power cord into the back of the DH-2000-CAL.

5. Plug the other end of the power cord into a standard outlet.

6. Turn the lamp power on using the black switch on the rear of the DH-2000-CAL.

7. Switch the individual lamps on using the buttons on the front of the DH-2000-CAL.

8. Allow the lamp to warm up for at least 20 minutes before proceeding.