2: DH-2000-CAL Specifications
8
000-10000-030-02-1209
Specifications
Specifications
Deuterium Lamp Criteria
Tungsten-Halogen Lamp
Criteria
Wavelength Range
220–400 nm
300–1050 nm
Current – Voltage – Stability
0,01 % / h
0.4% of voltage
Current – Voltage – Drift
0,01 % / h
Warm-Up Time
20 minutes
Lamp Voltage
Lamp Current
Ignition 580V / 20°
Operating 85 V / 0.3A
12V DC / 1.67A
Lamp Lifetime
1000 hours
Radiation Characteristics
Aperture 0.5mm NA26° (13°)
Focused
Performance Guaranteed
Temperature
5°C – 35°C
Humidity
5 - 95% without condensation at 40°
Internal Power Consumption
25 W
20 W
Total Power
Max. Power Consumption
100 Watt/190 Watt (Heating D-Lamp for 20sec)
Power Requirements:
European Version (prior to 4/2003)
USA Version (prior to 4/2003)
All units manufactured after 4/2003
230-240V 50/60 Hz
110-115V 50/60 Hz
90-240V 50/60 Hz
Markings / Directives
CE; VDI/VDE 0160; EN 61010
Weight
6 kg
Size
150 x 135 x 319 mm
- Apex 785 Raman (1 page)
- STS-UV (2 pages)
- TR2 Engineering Note (4 pages)
- SAD500 Communications and Control (19 pages)
- Red Tide USB650 Install (26 pages)
- Fiber Optic Termination Kit (6 pages)
- Transmissive pH Probe (10 pages)
- Remora (42 pages)
- PlasCalc (59 pages)
- Correcting Device Driver Issues (8 pages)
- ecoVis Krypton Light Source (16 pages)
- LPC-500CM (28 pages)
- HPX-2000 (24 pages)
- ADC1000-USB (27 pages)
- Torus Operating Instructions (30 pages)
- ADC2000-PCI (13 pages)
- Sensors for Real-Time Analysis (2 pages)
- IDRaman reader (2 pages)
- QE65 Pro (32 pages)
- Collimating Lenses (2 pages)
- D1000 (2 pages)
- IDRaman mini (2 pages)
- HR2000CG-UV-NIR (42 pages)
- Cool Red (1 page)
- HL-2000 (20 pages)
- XE-1 Xenon (6 pages)
- USB-ADP Serial Adapters (3 pages)
- CHEM2000 (36 pages)
- Deuterium-Halogen Calibration Light Source (19 pages)
- NeoFox Engineering Note (30 pages)
- QE Pro (2 pages)
- OOIBase32 (140 pages)
- AR-1 Argon (6 pages)
- pH Sensor Patches, Probes and Cuvettes (36 pages)
- SpecLine Offline Spectroscopy (60 pages)
- HL-2000-HP-232 (26 pages)
- External Triggering Options Instructions for Spectrometers with Firmware Version 3.0 and Above (16 pages)
- Breakout Box (10 pages)
- USB-ISS-UV_VIS (4 pages)
- ISS-UV_VIS (6 pages)
- OOIColor (14 pages)
- LS-1 Series (12 pages)
- Apex Install (24 pages)
- DH-2000 (34 pages)