Using the ADC1000-USB: Serial Mode

Saving Wavelength Calibration Coefficients: Serial Mode

If you are interfacing the S2000 Spectrometer or S1024DW Spectrometer to your PC via the serial
cable, you need to save the wavelength calibration coefficients to the .spec file that OOIBase32
accesses each time it opens.

Note: You cannot save the calibration coefficients to the EEPROM memory chip on the

ADC1000-USB when using the serial mode.

To save wavelength calibration coefficients using the serial mode:

1. Open OOIBase32 and select Spectrometer > Configure from the menu bar.

2. Choose

the

Wavelength Calibration tab to update the wavelength coefficients within

OOIBase32.

3. Enter in the values you acquired from Steps 5-6 in the Calibrating the Wavelength of the

Spectrometer section.

ADC1000-USB Operating Instructions

14

4. Click on the Apply button, then click the OK button.