Measuring discrete samples – Ocean Optics LPC-500CM User Manual

Page 20

background image

3: Operation

12

041-00000-010-02-201106

5. Flush the LWCC with de-ionized water or experimental buffer solution using a pump or a syringe

at a rate of 1mL/minute and observe the light intensity or absorbance baseline on the detector.
Continue flushing until the signal is stable. (See

Troubleshooting

in Chapter 4:

Maintenance

if

the signal does not stabilize appropriately.)

Measuring Discrete Samples

Discrete samples can be measured with the LWCC by introducing the sample with a syringe. Sample
volumes approximately 1.5 – 3 times the cell volume are necessary to fill the LWCC. When injecting a
sample or reference solution, apply consistent hand pressure to the syringe. Small variations in baseline
levels may result from using this method.

Procedure

1. Using fiber optic cables, connect the LWCC to a light source and a detector.

2. Clean the LWCC using the standard cleaning procedure described in Chapter 4:

Maintenance

.

3. Flush the LWCC with de-ionized water or experimental buffer solution using a syringe and

observe the light intensity or absorbance baseline on the detector until the signal is stable.

4. Introduce the sample slowly into the LWCC with a syringe using steady pressure to avoid

generating air bubbles.