Specifications – Ocean Optics Transmissive pH Probe User Manual
Page 10

pH Probe Instructions
10
PH-BCG-TRANS-01-0210
Performing a linear fit gives a line with pK equal to the y-intercept and slope equal to the slope. In the
example chart above, the new pK value would be 6.2977 and the new slope value would be 1.7248.
Specifications
Specification
Ocean Optics pH Probe Value
pH Range
5 – 9
Temperature Range
-5 – 80°C
Accuracy
<1% of reading across range
Resolution
0.02 pH
Response Time (t
90
)
30 seconds
Response time increases with decreasing ionic strength
Calibration Requirements (minimum) 3 buffers
Sterilization
Gamma, EtO, Autoclave
Chemical Compatibility
Aqueous, alcohols, some organic solvents, peroxides,
ammonia, sodium hypochlorite
Chemicals to Avoid
Concentrated acids
Drift (Continuous Stability)
1% per day
Discrete Stability (Lifetime)
50 uses or more, dispose when absorbance at
pH 11 < 0.1 (assumes pH 1 reference)
Storage Conditions
Dry or wet storage
- Apex 785 Raman (1 page)
- STS-UV (2 pages)
- TR2 Engineering Note (4 pages)
- SAD500 Communications and Control (19 pages)
- Red Tide USB650 Install (26 pages)
- Fiber Optic Termination Kit (6 pages)
- Remora (42 pages)
- PlasCalc (59 pages)
- Correcting Device Driver Issues (8 pages)
- ecoVis Krypton Light Source (16 pages)
- LPC-500CM (28 pages)
- HPX-2000 (24 pages)
- ADC1000-USB (27 pages)
- Torus Operating Instructions (30 pages)
- ADC2000-PCI (13 pages)
- Sensors for Real-Time Analysis (2 pages)
- IDRaman reader (2 pages)
- DH-2000-CAL (30 pages)
- QE65 Pro (32 pages)
- Collimating Lenses (2 pages)
- D1000 (2 pages)
- IDRaman mini (2 pages)
- HR2000CG-UV-NIR (42 pages)
- Cool Red (1 page)
- HL-2000 (20 pages)
- XE-1 Xenon (6 pages)
- USB-ADP Serial Adapters (3 pages)
- CHEM2000 (36 pages)
- Deuterium-Halogen Calibration Light Source (19 pages)
- NeoFox Engineering Note (30 pages)
- QE Pro (2 pages)
- OOIBase32 (140 pages)
- AR-1 Argon (6 pages)
- pH Sensor Patches, Probes and Cuvettes (36 pages)
- SpecLine Offline Spectroscopy (60 pages)
- HL-2000-HP-232 (26 pages)
- External Triggering Options Instructions for Spectrometers with Firmware Version 3.0 and Above (16 pages)
- Breakout Box (10 pages)
- USB-ISS-UV_VIS (4 pages)
- ISS-UV_VIS (6 pages)
- OOIColor (14 pages)
- LS-1 Series (12 pages)
- Apex Install (24 pages)
- DH-2000 (34 pages)