beautypg.com

Specifications – Ocean Optics Transmissive pH Probe User Manual

Page 10

background image

pH Probe Instructions

10

PH-BCG-TRANS-01-0210

Performing a linear fit gives a line with pK equal to the y-intercept and slope equal to the slope. In the
example chart above, the new pK value would be 6.2977 and the new slope value would be 1.7248.

Specifications

Specification

Ocean Optics pH Probe Value

pH Range

5 – 9

Temperature Range

-5 – 80°C

Accuracy

<1% of reading across range

Resolution

0.02 pH

Response Time (t

90

)

30 seconds
Response time increases with decreasing ionic strength

Calibration Requirements (minimum) 3 buffers

Sterilization

Gamma, EtO, Autoclave

Chemical Compatibility

Aqueous, alcohols, some organic solvents, peroxides,

ammonia, sodium hypochlorite

Chemicals to Avoid

Concentrated acids

Drift (Continuous Stability)

1% per day

Discrete Stability (Lifetime)

50 uses or more, dispose when absorbance at
pH 11 < 0.1 (assumes pH 1 reference)

Storage Conditions

Dry or wet storage