beautypg.com

VEGA VEGABAR 82 Profibus PA - Operating Instructions User Manual

Page 67

background image

67

11 Supplement

VEGABAR 82 • Profibus PA

45031-EN-131107

Nominal measuring range in bar/

kPa

Nominal measuring

range in psig

In the compensated

temperature range

In the non-compen-

sated temperature

range

0 … 1 bar/0 … 100 kPa

0 … 15 psig

< {0.1 % + 0.2 x

(0.5+0.5 x TD)}

< {0.1 % + 0.3 x

(0.5+0.5 x TD)}

0 … 2.5 bar/0 … 250 kPa

0 … 35 psig

-0.5 … 0.5 bar/-50 … 50 kPa

-7 … 7 psig

-0.2 … 0.2 bar/-20 … 20 kPa

-3 … 3 psig

< {0.15 % + 0.4 x

(0.5+0.5 x TD)}

< {0.15 % + 0.6 x

(0.5+0.5 x TD)}

0 … 0.4 bar/0 … 40 kPa

0 … 6 psig

>>>Long-term stability (according to DIN 16086 and IEC 60770-1)
Specifications refer to the set span. Turn down (TD) is the ratio: nominal measuring range/set span.
Long-term drift of the zero signal

Time pe-

riod

Measuring cell ø 28 mm Measuring cell ø 28 mm

Measuring range

0 … +0.025 bar/0 … +2.5 kPa

Measuring cell

ø 17.5 mm

One year

< 0.05 % x TD

< 0.1 % x TD

< 0.1 % x TD

Five years

< 0.1 % x TD

< 0.2 % x TD

< 0.2 % x TD

Ten years

< 0.2 % x TD

< 0.4 % x TD

< 0.4 % x TD

Long-term drift of the zero signal - Version climate-compensated

Nominal measuring range in bar/

kPa

Nominal meas-

uring range in

psig

Measuring cell ø 28 mm Measuring cell

ø 17.5 mm

0 … 0.4 bar/0 … 40 kPa

0 … 6 psig

< (1 % x TD)/year

< (1.5 % x TD)/year

-0.2 … 0.2 bar/-20 … 20 kPa

-3 … 3 psig

0 … 1 bar/0 … 100 kPa

0 … 15 psig

< (0.25 % x TD)/year

< (0.375 % x TD)/year

0 … 2.5 bar/0 … 250 kPa

0 … 35 psig

-1 … 0 bar/-100 … 0 kPa

-15 … 0 psig

-1 … 1.5 bar/-100 … 150 kPa

-15 … 25 psig

-0.5 … 0.5 bar/-50 … 50 kPa

-7 … 7 psig

0 … 10 bar/0 … 1000 kPa

0 … 150 psig

< (0.1 % x TD)/year

< (0.15 % x TD)/year

0 … 25 bar/0 … 2500 kPa

0 … 350 psig

0 … 60 bar/0 … 6000 kPa

0 … 900 psig

0 … 100 bar/0 … 6000 kPa

0 … 1450 psig

-1 … 10 bar/-100 … 1000 kPa

-15 … 150 psig

-1 … 25 bar/-100 … 2500 kPa

-15 … 350 psig

-1 … 60 bar/-100 … 6000 kPa

-15 … 900 psig