6 Set up with the display and adjustment module

VEGABAR 82 • Profibus PA

45031-EN-131107

Enter the requested parameters via the appropriate keys, save your

settings with [OK] and jump to the next menu item with the [ESC] and

the [->] key.

In this menu item, the adjustment units of the instrument are deter-

mined. The selection determines the displayed unit in the menu items

"Min. adjustment (zero)" and "Max. adjustment (span)".
Unit of measurement:

If the level should be adjusted in a height unit, the density of the me-

dium must also be entered later during the adjustment.
In addition, the temperature unit of the instrument is specified. The se-

lection determines the displayed unit in the menu items "Peak value,

measuring cell temperature", "Peak value, electronics temperature"

as well as "HART variables".
Temperature unit:

Enter the requested parameters via the appropriate keys, save your

settings with [OK] and jump to the next menu item with the [ESC] and

the [->] key.

Especially with chemical seal systems, the installation position of the

instrument can shift (offset) the measured value. Position correction

compensates this offset. In the process, the actual measured value

is taken over automatically. With relative pressure measuring cells a

manual offset can also be carried out.

If the actual measured value should be taken over as correction value

during automatic position correction, this value must not be influ-

enced by product coverage or static pressure.
With the manual position correction, the offset value can be deter-

mined by the user. Select for this purpose the function "Edit" and

enter the requested value.
Save your settings with [OK] and move with [ESC] and [->] to the

next menu item.

Setup - Units

Setup - Position correc-

tion