VEGA VEGAFLEX 82 4 … 20 mA_HART four-wire User Manual
Page 73

73
11 Supplement
VEGAFLEX 82 • 4 … 20 mA/HART - four-wire
41830-EN-130314
1
2
0°C
(32°F)
-40°C
(-104°F)
80°C
(176°F)
150°C
(302°F)
-40°C
(-104°F)
Fig. 43: Ambient temperature - process temperature, standard version
Cable version with ø 6 mm (0.236 in)
Cable version, PA coated with ø 11 mm (0.433 in)
1 Ambient temperature
2 Process temperature (depending on the seal material)
1
2
0°C
(32°F)
-40°C
(-104°F)
65°C
(149°F)
80°C
(176°F)
200°C
(392°F)
170°C
(338°F)
3
4
-20°C
(-68°F)
Fig. 44: Ambient temperature - process temperature, version with temperature adapter
Cable version with ø 6 mm (0.236 in)
Cable version, PA coated with ø 11 mm (0.433 in)
1 Ambient temperature
2 Process temperature (depending on the seal material)
3 Maximum permissible temperature - standard
4 Limited temperature range - plastic housing and stainless steel housing, electropolished
Vibration resistance
Ʋ Instrument housing
4 g at 5 … 200 Hz according to EN 60068-2-6 (vibration
with resonance)
Ʋ Rod probe
1 g with 5 … 200 Hz according EN 60068-2-6 (vibration
at resonance) with rod length 50 cm (19.69 in)
Shock resistance
Ʋ Instrument housing
100 g, 6 ms according to EN 60068-2-27 (mechanical
shock)