9 Diagnostics and service
VEGAFLEX 82 • 4 … 20 mA/HART - four-wire
41830-EN-130314
6. Cable: shift the cable into the gravity weight (according to the
drawing)
Plastic coated cable: remove coating according drawing to 70 mm
(2.76 in).
7. Cable: Fasten the cable with three pins, torque 20 Nm
(14.75 lbf in)
Cable ø 4: 7 Nm (5.16 lbf ft)
Cable ø 6, cable ø 8: 20 Nm (14.75 lbf ft)
8. Enter new probe length and then carry out a fresh adjustment
(see "Setup procedure, Carrying out min. adjustment - Carrying
out max. adjustment").
3
1
2
2
A
B
45 mm (1.77")
150 mm (5.91")
70 mm (2.76")
150 mm (5.91")
Fig. 33: Shortening the cable probe
A Gravity weight - cable ø 4 mm
B Gravity weight - cable ø 6 mm
1 Pins
2 Thread M12 for eye-bolt
3 Pins
9.7 Software update
The following components are required to update the sensor soft-
ware:
•
Sensor
•
Voltage supply
•
Interface adapter VEGACONNECT 4
•
PC with PACTware
•
Current sensor software as file
You can find the actual sensor software as well as detailed infor-
mation of the procedure under "www.vega.com/downloads" and
"Software".
You can find information about the installation in the download file.