11 supplement – VEGA VEGAFLEX 86 Foundation Fieldbus User Manual
Page 85
85
11 Supplement
VEGAFLEX 86 • Foundation Fieldbus
44233-EN-130910
Process temperature
-196 … +450 °C (-321 … +842 °F)
The measurement error from the process conditions is in
the specified pressure and temperature range of below
1 %.
1
2
0°C
(32°F)
-40°C
(-104°F)
80°C
(176°F)
450°C
(842°F)
220°C
(428°F)
3
4
-196°C
(-384.8°F)
-20°C
(-68°F)
-110°C
(-230°F)
35°C
(95°F)
Fig. 63: Ambient temperature - process temperature (version -196 … +450 °C/-321 … +842 °F)
1 Ambient temperature
2 Process temperature (depending on the seal material)
3 Maximum permissible temperature - standard
4 Limited temperature range - plastic housing and stainless steel housing, electropolished
2
1
-100°C
(-148°F)
-200°C
(-328°F)
100 bar
(1450 psig)
200 bar
(2900 psig)
300 bar
(4350 psig)
400 bar
(5800 psig)
380 bar
(5510 psig)
350 bar
(5070 psig)
0°C
(32°F)
200°C
(392°F)
100°C
(212°F)
400°C
(752°F)
300°C
(572°F)
320 bar
(4640 psig)
450°C
(842°F)
Fig. 64: Process pressure - process temperature (version -196 … +450 °C/-321 … +842 °F)
1 Process temperature (depending on the seal material)
2 Process pressure
Vibration resistance
Ʋ Instrument housing
4 g at 5 … 200 Hz according to EN 60068-2-6 (vibration
with resonance)
Ʋ Rod probe
1 g with 5 … 200 Hz according EN 60068-2-6 (vibration
at resonance) with rod length 50 cm (19.69 in)