beautypg.com

VEGA MINITRAC 31 Foundation Fieldbus User Manual

Page 64

background image

64

9 Diagnostics and service

MINITRAC 31 • Foundation Fieldbus

41782-EN-131119

Code
Text mes-

sage

Cause

Rectification

F036
Faulty pro-

gram memory

– Error during software

update

– Repeat software update

– Exchanging the electronics

F037
RAM hard-

ware error

– Error in RAM

– Restart instrument

– Exchanging the electronics

F038
Slave signals

failure

– Connection cable to the

Slave instrument interrupted

– Instrument not defined as

Slave instrument

– Check the connection cable

to the Slave instrument

– Define instrument as Slave

F040
Hardware

error

– Sensor defective

– Restart instrument

– Exchanging the electronics

F041
Photomultipli-

er error

– Error in the measured value

recording

– Restart instrument

– Exchanging the electronics

F052
Faulty config-

uration

– Invalid parameter adjust-

ment

– Carry out a reset

F066
Faulty adjust-

ment

– Adjustment not yet carried

out

– Error during adjustment or

when entering the lineariza-

tion table

– Carrying out adjustment

– Carry out linearization

F068
Count rate too

high

– Faulty instrument settings

– Carry out a reset

F072
Limit exceed-

ed

– Faulty instrument settings

– Carry out a reset

F073
Error real val-

ue correction

– Faulty real value correction

– Carry out real value correc-

tion again

F080
System error

– Instrument error

– Restart instrument

– Call our service

F086
Communica-

tion error

– Error in the Fieldbus com-

munication

– Restart instrument

– Call our service

F114
Error real time

clock

– Discharge accumulator

– Readjust real time clock

F120
Filter time

error

– Faulty or missing instrument

adjustment

– Carrying out adjustment