VEGA VEGAFLEX 86 Modbus and Levelmaster protocol User Manual
Page 88
88
11 Supplement
VEGAFLEX 86 • Modbus and Levelmaster protocol
41848-EN-130314
Parameter
Configurable Values
Default Value
Baud Rate
1200, 2400, 4800, 9600, 19200
9600
Start Bits
1
1
Data Bits
7, 8
8
Parity
None, Odd, Even
None
Stop Bits
1, 2
1
Address range Levelmaster
32
32
The Levelmaster commands are based on the following syntax:
•
Capital letters are at the beginning of certain data fields
•
Small letters stand for data fields
•
All commands are terminated with "
•
All commands start with "Uuu", whereby "uu" stands for the address (00-31)
•
"*" can be used as a joker for any position in the address. The sensor always converts this in
its address. In case of more than one sensor, the joker must not be used, because otherwise
several slaves will answer
•
Commands that modify the instrument return the command with "OK“. "EE-ERROR" replaces
"OK" if there was a problem changing the configuration
Report Level (and Temperature)
Request:
Parameter
Length
Code/Data
Report Level (and Temperature)
4 characters ASCII
Uuu?
Response:
Parameter
Length
Code/Data
Report Level (and Temperature)
24 characters ASCII
UuuDlll.llFtttEeeeeWwwww
uu = Address
lll.ll = PV in inches
ttt = Temperature in Fahrenheit
eeee = Error number (0 no error, 1
level data not readable)
wwww = Warning number (0 no
warning)
Level in inches is repeated for sensors with 2 levels. PV value is transmitted as first measured value,
SV as second measured value.
PV, SV, TV and QV can be adjusted via the sensor DTM.
Report Unit Number
Request:
Parameter
Length
Code/Data
Report Unit Number
5 characters ASCII
U**N?