beautypg.com

VEGA VEGAFLEX 81 HART - integrated accumulator User Manual

Page 43

background image

43

6 Set up with the display and adjustment module

VEGAFLEX 81 • HART - integrated accumulator

44640-EN-130430

With a reset, certain parameter adjustments carried out by the user

are reset.

The following reset functions are available:
Delivery status: Restoring the parameter settings at the time of ship-

ment from the factory incl. the order-specific settings. A created false

signal suppression, user-programmable linearization curve as well as

the measured value memory will be deleted.
Basic settings: Resetting the parameter settings incl. special pa-

rameters to the default values of the respective instrument. A created

false signal suppression, user programmable linearization curve as

well as the measured value memory will be deleted.
The following table shows the default values of the instrument. De-

pending on the instrument version or application, all menu items may

not be available or some may be differently assigned:

Setup

Menu item

Default value

Block operation

Released

Measurement loop name Sensor
Units

Distance unit: mm
Temperature unit: °C

Probe length

Length of the probe Ex factory

Medium type

Liquid

Applications

Level, vessel

Medium, dielectric con-

stant

Water-based, > 10

Superimposed gas

phase

Yes

Dielectric figure, upper

medium (TS)

1.5

Tube inner diameter

200 mm

Max. adjustment - Level 100 %
Max. adjustment - Level Distance: 0.000 m(d) - note blocking distances
Min. adjustment - Level

0 %

Additional adjustments

- Reset