7 Set up with the display and adjustment module

MINITRAC 31 • 4 … 20 mA/HART - four-wire

43389-EN-131119

Select "Linearization - New" to enter the first point.

Select "Determine count rate" to enter the first point.

The determination of the actual count rate lasts 2 minutes. After the

count rate has been determined, you can accept the value (ct/s).
The count rate is stated in ct/s. This is the number of counts per sec-

ond, i.e. the measured radioactive radiation dose actually reaching

the sensor.

Enter now the corresponding level (m).
By doing so, you assign a respective level to the actual count rate.

Accept the value pair with "OK".
Depending if you have started with full or empty vessel, you have to

continue emptying or filling the vessel.
Also carry out such a linearization with several different filling heights

if you have a linear vessel.
By doing so, you can influence the measurement reliability of the

sensor. The more linearization points you enter and the bigger the

difference between the first and the last linearization point, the more

reliable the measurement will be.
If you have a non-linear vessel (e.g. spherical tank), you have to carry

out this adjustment with as many points as possible.
A maximum of 32 linearisation points is possible.
Note:

If you cannot fill the vessel during the adjustment procedure to at least

60 % of the max. level, it is possible to carry out the full adjustment

by switching off the radiation. The switched-off radiation simulates a

100 % filling.