Nominal measuring range in bar/kPa

Nominal measuring
range in psig

Long-term drift of the zero signal

-0.1 … 0.1

bar/-10 … 10 kPa

-1.5 … 1.5

psig

< (1 %

x TD)/year

-0.2 … 0.2

bar/-20 … 20 kPa

-3 … 3

psig

< (1 %

x TD)/year

-0.5 … 0.5

bar/-50 … 50 kPa

-7 … 7

psig

< (0.25 %

x TD)/year

Ambient conditions

A

mbient, storage and transport temperature

-40 … +80 °C (-40 … +176 °F)

Process conditions

T

he specifications to the pressure stage and the product temperature are used as an overview.

T

he specifications of the type label are applicable.

P

ressure stage, process fitting

-

T

hread 316L, Hastelloy

PN

160

-

T

hread PVDF

PN

10

-

S

mall flange DN 10 316L

PN

1.5

P

roduct temperature depending on the measuring cell seal

11)

-

FKM (A+P 70, 16.06)

-40 … +120 °C (-40 … +248 °F)

-

FFKM (K

alrez 6375)

-20 … +120 °C (-4 … +248 °F)

-

FFKM (C

hemraz 535)

-30 … +120 °C (-22 … +248 °F)

-

EPDM (A+P 75.5/KW75F)

-40 … +120 °C (-40 … +248 °F)

V

ibration resistance

mechanical vibrations with 4 g and 5 … 100 Hz

12)

S

hock resistance

A

cceleration 100 g/6 ms

13)

50

VEGABAR

52 • 4 … 20

mA/HART - climate compensated

10 S

upplement

35075

-EN