VEGA VEGABAR 52 - climate compensated User Manual
Page 49
Nominal measuring range in
bar/kPa
Nominal
measuring
range in psig
Average temperature coeffi-
cient of the zero signal9)
Average temperature coeffi-
cient of the zero signal10)
0 … 10
bar/0 … 1000 kPa
0 … 150
psig < (0.02 + 0.03 x TD)%/10K
< (0.04 + 0.06
x TD)%/10K
0 … 25
bar/0 … 2500 kPa
0 … 350
psig < (0.02 + 0.03 x TD)%/10K
< (0.04 + 0.06
x TD)%/10K
0 … 60
bar/0 … 6000 kPa
0 … 900
psig < (0.02 + 0.03 x TD)%/10K
< (0.04 + 0.06
x TD)%/10K
-1 … 0
bar/-100 … 0 kPa
-15 … 0
psig
< (0.02 + 0.03
x TD)%/10K
< (0.04 + 0.06
x TD)%/10K
-1 … 1.5
bar/-100 … 150 kPa -15 … 25 psig < (0.02 + 0.03 x TD)%/10K
< (0.04 + 0.06
x TD)%/10K
-1 … 5
bar/-100 … 500 kPa
-15 … 70
psig < (0.02 + 0.03 x TD)%/10K
< (0.04 + 0.06
x TD)%/10K
-1 … 10
bar/-100 … 1000 kPa -15 … 150 psig < (0.02 + 0.03 x TD)%/10K
< (0.04 + 0.06
x TD)%/10K
-1 … 25
bar/-100 … 2500 kPa -15 … 350 psig < (0.02 + 0.03 x TD)%/10K
< (0.04 + 0.06
x TD)%/10K
-1 … 60
bar/-100 … 6000 kPa -15 … 900 psig < (0.02 + 0.03 x TD)%/10K
< (0.04 + 0.06
x TD)%/10K
-0.1 … 0.1
bar/-10 … 10 kPa
-1.5 … 1.5
psig < (0.05 + 0.05 x TD)%/10K
< (0.05 + 0.1
x TD)%/10K
-0.2 … 0.2
bar/-20 … 20 kPa
-3 … 3
psig
< (0.05 + 0.05
x TD)%/10K
< (0.05 + 0.1
x TD)%/10K
-0.5 … 0.5
bar/-50 … 50 kPa
-7 … 7
psig
< (0.05 + 0.05
x TD)%/10K
< (0.02 + 0.1
x TD)%/10K
Thermal change of the current output
A
pplies also to the analogue 4 … 20 mA current output and refers to the set span.
T
hermal change, current output
<
0.15 %
at -40 … +80 °C (-40 … +176 °F)
Long-term stability (similar to DIN 16086, DINV 19259-1 and IEC 60770-1)
A
pplies to digital HART interface as well as to analogue current output 4 … 20 mA. Specifications
refer to the set span. Turn down (TD) is the relation nominal measuring range/set span.
Nominal measuring range in bar/kPa
Nominal measuring
range in psig
Long-term drift of the zero signal
0 … 0.2
bar/0 … 20 kPa
0 … 3
psig
< (1 %
x TD)/year
0 … 0.4
bar/0 … 40 kPa
0 … 6
psig
< (1 %
x TD)/year
0 … 1
bar/0 … 100 kPa
0 … 15
psig
< (0.25 %
x TD)/year
0 … 2.5
bar/0 … 250 kPa
0 … 35
psig
< (0.25 %
x TD)/year
0 … 5
bar/0 … 500 kPa
0 … 70
psig
< (0.1 %
x TD)/year
0 … 10
bar/0 … 1000 kPa
0 … 150
psig
< (0.1 %
x TD)/year
0 … 25
bar/0 … 2500 kPa
0 … 350
psig
< (0.1 %
x TD)/year
0 … 60
bar/0 … 6000 kPa
0 … 900
psig
< (0.1 %
x TD)/year
-1 … 0
bar/-100 … 0 kPa
-15 … 0
psig
< (0.25 %
x TD)/year
-1 … 1.5
bar/-100 … 150 kPa
-15 … 25
psig
< (0.25 %
x TD)/year
-1 … 5
bar/-100 … 500 kPa
-15 … 70
psig
< (0.1 %
x TD)/year
-1 … 10
bar/-100 … 1000 kPa
-15 … 150
psig
< (0.1 %
x TD)/year
-1 … 25
bar/-100 … 2500 kPa
-15 … 350
psig
< (0.1 %
x TD)/year
-1 … 60
bar/-100 … 6000 kPa
-15 … 900
psig
< (0.1 %
x TD)/year
VEGABAR
52 • 4 … 20
mA/HART - climate compensated
49
10 S
upplement
35075
-
EN
-081209