VEGA VEGABAR 52 - climate compensated User Manual
Page 32
Menu section
Function
Reset value
B
asic settings
U
nit of measurement
bar
T
emperature unit
°C
Z
ero/Min. adjustment
M
easuring range begin
S
pan/Max. adjustment
M
easuring range end
D
ensity
1
kg/l
D
ensity unit
kg/l
D
amping
1
s
L
inearisation
linear
S
ensor-TAG
S
ensor
D
isplay
D
isplayed value 1
bar
D
isplayed value 2
%
D
isplay unit
V
olume/l
S
caling
0.00
to 100.0
D
ecimal point indication
8888.8
S
ervice
C
urrent output - characteristics
4 … 20
mA
C
urrent output - failure
<
3.6
mA
C
urrent output - min. current
3.8
mA
C
urrent output - max. current
20.5
mA
T
he values of the following menu items are not reset with "Reset:
Menu section
Function
Reset value
B
asic settings
P
osition correction
no reset
D
isplay
L
ighting
no reset
S
ervice
SIL
no reset
L
anguage
no reset
HART
mode
no reset
A
pplication
no reset
Factory setting
L
ike basic adjustment, furthermore special parameters are reset to
default values.
5)
Pointer
T
he min. and max. temperature or pressure values are each reset to
the actual value.
A
dditional adjustment and diagnosis options such as e.g. scaling,
simulation or trend curve presentation are shown in the following menu
schematic. You will find a detailed description of these menu items in
the operating instructions manual "Indicating and adjustment module".
5)
S
pecial parameters are parameters which are set customer-specifically on
the service level with the adjustment software PACTware.
Optional settings
32
VEGABAR
52 • 4 … 20
mA/HART - climate compensated
6 S
et up with the indicating and adjustment module PLICSCOM
35075
-EN
-081209