beautypg.com

VEGA VEGADIF 65 4 … 20 mA_HART User Manual

Page 77

background image

77

11 Supplement

VEGADIF 65 • 4 … 20 mA/HART

36128-EN-130417

Max. torque screws mounting strap

30 Nm

Max. torque screws socket external

housing

5 Nm (3.688 lbf ft)

Weight approx.

4.2 … 4.5 kg (9.26 … 9.92 lbs), depending on process

fitting

Output variable
Output signal

4 … 20 mA/HART

HART output values

Ʋ HART value (Primary Value)

Process pressure

Ʋ HART value (Secondary Value)

Temperature

Signal resolution

1.6 µA

Failure signal current output (adjustable) mA-value unchanged 20.5 mA, 22 mA, < 3.6 mA
Max. output current

22 mA

Load

see load diagram under Power supply

Met NAMUR recommendation

NE 43

Dynamic behaviour output
Run-up time

≤ 20 s

I

63 %

100 %

t

t

1

t

2

Fig. 59: Presentation of the dead time t

1

and the time constant t

2

The following specified total dead time applies to the 4 … 20 mA current output:

Version, nominal measuring range

Dead time t

1

Time constant t

2

Basic version 10 bar and 30 bar

100 ms

250 ms

Basic version 100 mbar

100 ms

180 ms

Basic version 500 mbar

100 ms

180 ms

Basic version, 3 bar

100 ms

180 ms

Basic version 16 bar and 40 bar

100 ms

180 ms

Chemical seal version, all nominal measuring ranges

Dependent on the chemi-

cal seal

Dependent on the chemi-

cal seal

Damping (63 % of the input variable)

0 … 999 s, adjustable