3 setup requirements – VEGA VEGALOG 571 CPU and carrier User Manual
Page 22
VEGACOM 558
The communication card VEGACOM 558 is an interface
converter for Ethernet connection via TCP/IP. Hence the
parameter adjustment of VEGALOG as well as reading out of
measured values from any network PC is possible. It can be
used as webserver and provides the measured values as
HTML pages to all network participants. E-mails can be sent
also with the actual measured values or event messages.
VEGACOM 558 can be also used for connection to Visual
VEGA. Here all measured values and trend curves can be
enquired via Ethernet and displayed on any PC.
VEGACOM 557
The VEGACOM 557 communication card is an interface
converter (Gateway) for conversion of VEGA-specific data
formats into standard protocols. It is used for connection of the
VEGALOG system to a PLC or a PLC system. The
communication card is available for the following protocols:
l
Siemens S5 (3964R procedure)
l
Modbus (RTU and ASCII)
l
Interbus S
l
Profibus FMS
l
Profibus DP
l
VEGA-ASCII
The communication card can also be used to provide data of
VEGALOG via RS232 to the visualisation software Visual
VEGA. For this case, the communication card is also available
without interface function.
Adapter print
With the adapter card VEGACOM 557 AP, data of standard
protocols (Profibus, Interbus etc.) which are normally only
available on the rear of the carrier, are led to the front plate.
VEGASTAB 593
The 19" power supply unit powers all cards of VEGALOG with
24 V DC. The sensors are also powered via the active inputs of
the input cards. The versions VEGASTAB 593-60 (24 V, 45 W)
and VEGASTAB 593 (24 V, 120 W) are available.
6.3 Setup requirements
The adjustment software PACTware™ with the corresponding
DTMs is installed under Windows™ installiert and enables
easy configuration of measuring systems as well as parameter
adjustment of connected VEGA sensors. For this purpose,
Communication cards
Power supply card
22
VEGALOG 571 - CPU and carrier
Set up