8 Maintenance and fault rectification
PLICSRADIO T62 • Wireless emitting unit (multiple channel)
32624-EN-130228
Error code Cause
Rectification
E034
EEPROM CRC error
– Switch the instrument off and on
– Carry out a reset
– Send instrument for repair
E035
ROM CRC error
– Switch the instrument off and on
– Carry out a reset
– Send instrument for repair
E036
Instrument software
not executable (dur-
ing software update
or after failed update)
– Wait until software update is finished
– Carry out another software update
E042
Hardware error with
selfcheck
– Send instrument for repair
E052
Double assignment
input 3 (sensor with
HART addr. 3 and
4 … 20 mA input con-
nected)
– Remove sensor with HART addr. 3 or
4 … 20 mA input
E053
Sensor measuring
range not read cor-
rectly
– Communication error: Check sensor
cable and screening
E086
Error communication
hardware (initiali-
sation of the radio
module failed)
– Initialisation is carried out automati-
cally. If the error exists permanently,
send instrument for repair
Depending on the reason for the fault and the measures taken, the
steps described in chapter "Set up" may have to be carried out again.
8.3 Instrument repair
If a repair is necessary, please proceed as follows:
You can download a return form (23 KB) from our Internet homepage
www.vega.com under: "Downloads - Forms and certificates - Repair
form".
By doing this you help us carry out the repair quickly and without hav-
ing to call back for needed information.
•
Print and fill out one form per instrument
•
Clean the instrument and pack it damage-proof
•
Attach the completed form and, if need be, also a safety data
sheet outside on the packaging
•
Please ask the agency serving you for the address of your return
shipment. You can find the respective contact data on our website
www.vega.com under: "Company - VEGA worldwide"