VEGA VEGAPULS 63 (≥ 2.0.0 - ≤ 3.8) Foundation Fieldbus User Manual
Page 53
-
TFM-PTFE hygienically encapsulated,
process seal FKM (Viton)
-25 … +130 °C (-13 … +266 °F)
-
TFM-PTFE hygienically encapsulated,
process seal EPDM
-40 … +130 °C (-40 … +266 °F)
Vibration resistance
mechanical vibrations with 4 g and 5 … 100 Hz
10)
Electromechanical data - version IP 66/IP 67 and IP 66/IP 68; 0.2 bar
Cable entry/plug
11)
-
Single chamber housing
l
1 x cable gland M20 x 1.5 (cable: ø 5 … 9 mm),
1 x blind stopper M20 x 1.5
or:
l
1 x closing cap M20 x 1.5; 1 x blind stopper
M
20 x 1.5
or:
l
1 x closing cap ½ NPT, 1 x blind plug ½ NPT
or:
l
1 x plug (depending on the version), 1 x blind
stopper M20 x 1.5
-
Double chamber housing
l
1 x cable entry M20 x 1.5 (cable: ø 5 … 9 mm),
1 x blind stopper M20 x 1.5; 1 x blind stopper
M
16 x 1.5 or optionally available with 1 x plug
M
12 x 1 for external indicating and adjustment
unit
or:
l
1 x closing cap ½ NPT, 1 x blind stopper
½ NPT
, 1 x blind stopper M16 x 1.5 or
optionally 1 x plug M12 x 1 for external
indicating and adjustment unit
or:
l
1 x plug (depending on the version), 1 x blind
stopper M20 x 1.5; 1 x blind stopper M16 x 1.5
or optionally available with 1 x plug M12 x 1 for
external indicating and adjustment unit
Spring-loaded terminals for wire cross-sec-
tion
< 2
.5 mm² (AWG 14)
Electromechanical data - version IP 66/IP 68 (1 bar)
Cable entry
-
Single chamber housing
1 x IP 68 cable gland M20 x 1.5; 1 x blind stopper
M
20 x 1.5
-
Double chamber housing
1 x IP 68 cable gland M20 x 1.5; 1 x blind stopper
M
20 x 1.5; 1 x blind stopper M16 x 1.5
10)
Tested according to the guidelines of German Lloyd, GL directive 2.
11)
Depending on the version M12 x 1, according to DIN 43650, Harting, 7/
8" FF.
VEGAPULS
63 • Foundation Fieldbus
53
10 Annex
28451
-
EN
-120530