2 parameter adjustment with pactware – VEGA VEGAMET 625 User Manual
Page 37
37
7 Setup with PACTware
VEGAMET 625 • Double channel HART
28970-EN-130701
3
4
1
2
Fig. 9: Connection of the PC via RS232
1 RS232 interface of the PC
2 RS232 interlink cable (article no. LOG571.17347)
3 RS232 modem connection cable (in the scope of delivery)
4 RS232 interface (RJ45 plug connection)
5 4 3 2 1
6
8 7
9
1
8
2
1
3
RXD
CTS
TXD
DTR
GND
RTS
4
2
3
1
5
6
2
8
3
4
5
7
Fig. 10: Terminal assignment of the RS232 modem connection cable
1 Name of the interface cable
2 Assignment of the RJ45 plug (view contact side)
3 Assignment of the RS232 plug (view soldering side)
7.2 Parameter adjustment with PACTware
As an alternative to the integrated display and adjustment unit, the
adjustment can be also carried out via a Windows PC. For this, the
configuration software PACTware and a suitable instrument driver
(DTM) according to the FDT standard are required. The current PACT-
ware version as well as all available DTMs are compiled in a DTM
Collection. Furthermore, the DTMs can be integrated into other frame
applications compliant with the FDT standard.
Note:
To ensure that all instrument functions are supported, you should
always use the latest DTM Collection. Furthermore, not all described
functions are included in older firmware versions. You can download
Assignment RS232 mo-
dem connection cable
Prerequisites