beautypg.com

3 product description – VEGA VEGASON 61 Foundation Fieldbus User Manual

Page 7

background image

7

3 Product description

VEGASON 61 • Foundation Fieldbus

28790-EN-130417

3 Product description

3.1 Configuration

The scope of delivery encompasses:

VEGASON 61 ultrasonic sensor

Documentation

– this operating instructions manual

– Operating instructions manual 27835 "Display and adjustment

module PLICSCOM" (optional)

– Supplementary instructions manual 31708 "Heating for display

and adjustment module" (optional)

– Supplementary instructions manual "Plug connector" (optional)

– Ex-specific "Safety instructions" (with Ex versions)

– if necessary, further certificates

The VEGASON 61 consists of the components:

Transducer with integrated temperature sensor

Housing with electronics, optionally available with plug connector

Housing cover, optionally available with display and adjustment

module PLICSCOM

The components are available in different versions.

1

2

3

Fig. 1: VEGASON 61 - version with plastic housing
1 Housing cover with integrated PLICSCOM (optional)

2 Housing with electronics, optionally available with plug connector

3 Process fitting with transducer

3.2 Principle of operation

VEGASON 61 is an ultrasonic sensor for continuous level measure-

ment. It is suitable for liquids and solids in virtually all industries,

particularly in the water and waste water industry.

The transducer of the ultrasonic sensor transmits short ultrasonic

pulses to the measured product. These pulses are reflected by

Scope of delivery

Constituent parts

Application area

Functional principle