background image

10

4 Mounting

VEGASON 61 • Foundation Fieldbus

28790-EN-130417

4 Mounting

4.1 General instructions

Make sure that all parts of the instrument coming in direct contact

with the process, especially the sensor element, process seal and

process fitting, are suitable for the existing process conditions, such

as process pressure, process temperature as well as the chemical

properties of the medium.
You can find the specifications in chapter "Technical data" and on the

nameplate.

Use the recommended cables (see chapter "Connecting to power

supply") and tighten the cable gland.
You can give your instrument additional protection against moisture

penetration by leading the connection cable downward in front of the

cable entry. Rain and condensation water can thus drain off. This ap-

plies mainly to outdoor mounting as well as installation in areas where

high humidity is expected (e.g. through cleaning processes) or on

cooled or heated vessels.

Fig. 2: Measures against moisture penetration

The reference plane for the measuring range is the lower edge of the

transducer.
Make sure that a minimum distance from the reference plane - the

so-called dead band, in which measurement is not possible - is

maintained. The exact value of the dead band is stated in chapter

"Technical data".

Suitability for the process

conditions

Moisture

Measuring range