10 supplement, 1 technical data – VEGA VEGASON 61 Foundation Fieldbus User Manual
Page 41
41
10 Supplement
VEGASON 61 • Foundation Fieldbus
28790-EN-130417
10 Supplement
10.1 Technical data
General data
Materials, wetted parts
Ʋ Process fitting, transducer
PVDF
Ʋ Seal between transducer and process
fitting
EPDM, FKM (Viton)
Materials, non-wetted parts
Ʋ Housing
Plastic PBT (polyester), Alu die-casting powder-coated,
316L
Ʋ Seal between housing and housing
cover
NBR (stainless steel housing), silicone (Alu/plastic hous-
ing)
Ʋ Inspection window in housing cover
Polycarbonate
Ʋ Ground terminal
316Ti/316L
Weight
1.8 … 4 kg (4 … 8.8 lbs), depending on the process
fitting and housing
Max. torque mounting boss
25 Nm
Output variable
Output
Ʋ Signal
digital output signal, Foundation Fieldbus protocol
Ʋ Physical layer
according to IEC 61158-2
Cycle time
min. 1 s (dependent on the parameter setting)
Ʋ Damping (63 % of the input variable) 0 … 999 s, adjustable
Ʋ Met NAMUR recommendation
NE 43
Channel Numbers
Ʋ Channel 1
Primary value
Ʋ Channel 2
Secondary value 1
Ʋ Channel 3
Secondary value 2
Transmission rate
31.25 Kbit/s
Current value
10 mA, ±0.5 mA
Resolution, digital
> 1 mm (0.039 in)
Input variable
Measured variable
distance between lower edge of the transducer and
product surface
Measuring range
Ʋ Liquids
up to 5 m (16.4 ft)
Ʋ Bulk solids
up to 2 m (6.562 ft)
Dead band
0.25 m (0.82 ft)
Reference conditions to measuring accuracy (according to DIN EN 60770-1)
Reference conditions according to DIN EN 61298-1