beautypg.com

VEGA VEGALOG 571 User Manual

Page 12

background image

12

VEGALOG 571

Product description

Electrical separating measures

Galvanic isolation
acc. to VDE 0106, part 1

between power supply, LOGBUS-connection

and meas. data inputs

- reference voltage

250 V

Common reference potential

between the individual meas. data inputs (GND)

Indicating elements

LED in front plate

green (on): mains voltage on
red: fault signal

EV-card

Meas. data inputs

Number

15 inputs

Data transmission

digital (VBUS)

Sensors

hydrostatic pressure transmitters
ultrasonic sensors
radar sensors

Sensor supply voltage

3 x 24 V DC, 1 A, short circuit proof

(for 5 sensors each)

Connection line

2-wire (screening recommended)

Line length

max. 1000 m

1)

Resistance per conductor

max. 15

1)

Interfaces

To LOGBUS

via LOGBUS-plug on the module card

Electrical separating measures

Galvanic isolation
acc. to VDE 0106, part 1

between power supply, LOGBUS-connection

and meas. data inputs

- reference voltage

250 V

Common potential

between VBUS-meas. data inputs

Indicating elements

LED in front facia

green (on): operating voltage on
red: fault signal

1)

The values are valid for a connected sensor. Note reduction acc. to “Connection instructions for VBUS-