VEGA Conductive electrodes User Manual
Page 10

10
Conductive measuring system
Product description
Measuring data input, Ex-technical data
Signal circuits
in classification intrinsic safety EEx ia IIC
Max. values
- U
O
10,6 V
- I
K
8 mA
- P
20 mW
Characteristics
linear
Inner effective inductance
negligible
Inner effective capacitance
negligible
Max. permissible outer inductance
5 mH
Max. permissible outer capacitance
570 nF
The intrinsically safe circuits are reliably galvanically separated from the not-intrinsically
safe circuits up to a peak value of the nominal voltage of 375 V.
The intrinsically safe circuits of channel 1 and channel 2 are reliably galvanically isolated.
Relay outputs
Number
2
Mode
overfill protection (A) and dry run protection (B)
changeover for each output (channel) separately
Contact
1 spdt per output
Contact material
AgCdO and Au plated
Turn-on voltage
min. 10 mV
max. 250 V AC or 250 V DC
Switching current
min. 10 µA
max. 3 A AC or 1 A DC
Breaking capacity
max. 750 VA or 54 W
Transistor outputs
Number
2 (synchronously switching with relay outputs)
Galvanic isolation
floating
Turn-on voltage
U
B
= max. 36 V DC
Switching current
I
B
= max. 60 mA
Voltage loss on the transistor
U
CE
= approx. 1,5 V at I
B
= 60 mA
Blocking current
I
O
< 10 µA
Integration time
Fixed value
approx. 500 ms
Ambient conditions
Permissible ambient temperature
-20°C … +60°C
Storage and transport temperature
-40°C … +70°C