GE Healthcare Avance User Manual

Page 7

background image

System storage

Temperature:

-25° to 60°C/-13° to 140°F

Humidity:

10 to 95% relative humidity
(non-condensing) per IEC 68-2-3

Altitude:

-440 to 5860 m/
375 to 800 mmHg

Oxygen cell storage:

-15° to 50°C/5° to 122°F
10 to 95% relative humidity
500 to 800 mmHg

Electromagnetic compatibility

Immunity:

Complies with all requirements
of EN 60601-1-2

Emissions:

CISPR 11 group 1 class B

Approvals:

UL 2601-1,
CSA C22.2 #601.1,
EN/IEC 60601-1,
CE 0197, EN 740

Breathing Circuit Specifications

Operational modes

Breathing circuit is circle mode; SCGO option converts to
open circuit mode

Carbon dioxide absorbent canister

Absorbent capacity:

800 g

Integrated expiratory limb water reservoír

Ports and connectors

Exhalation:

22 mm OD ISO
15 mm ID taper

Inhalation:

22 mm OD ISO
15 mm ID taper

Bag port:

22 mm OD

Bag-to-Ventilator switch

Type:

Bi-stable

Control:

Controls ventilator and direction
of breathing gas within the circuit

Integrated Adjustable Pressure Limiting (APL) valve

Range:

0.8 to 70 cm H

2

O

Tactile knob indication at:

30 cm H

2

O and above

Adjustment range
of rotation:

0.8 to 30 cm H

2

O (0 to 230°)

30 to 70 cm H

2

O (230 to 330°)

Materials

All materials in contact with exhaled patient gases are
autoclavable, except disposable flow sensors, O

2

cell,

and M-CAiOVX or ECAiOVX module. (Autoclavable flow
sensors optional)

All materials in contact with patient gas are free of natural
rubber latex.

Breathing circuit parameters

Compliance:

Bag mode:

1.82 mL/cm H

2

O

Mechanical mode:

Automatically
compensates
for compression
losses within the
absorber and
bellows assembly

Total circuit volume: 2.7 L Vent Mode

1.2 L Bag Mode

Note: Includes Absorber volume

Expiratory
resistance:

P

exp

P

exp

Bag Mode

Vent Mode

Flow rate

Pressure drop

Pressure drop

10 L/min

0.78 cm H

2

O

0.77 cm H

2

O

30 L/min

1.59 cm H

2

O

1.71 cm H

2

O

60 L/min

3.48 cm H

2

O

3.88 cm H

2

O

Note: With patient circuit and wye piece add +0.89 cm H

2

O

Anesthetic gas scavenging

AGSS Type

Hospital extract

Machine

system required

connection

High vacuum, low
flow with indicator:

High vacuum

DISS evac

36 L/min @ 12 in Hg
(305 mmHg)

High vacuum,
variable flow with
bag indicator:

High vacuum

DISS evac

30 L/min extract
flow @ 12 in Hg
(305 mmHg)

Passive:

Passive or external

30 mm/1.2 in

active system

M ISO taper

with air break

7

M1042385-2.qxd:M1042385-1 19.11.2008 10:48 Sivu7