O
2
waveform
Measurement range:
0 to 100 %
Accuracy:
±2 vol %*
Datex-Ohmeda differential paramagnetic sensor
Adjustable low and high alarm limits for FiO
2
and EtO
2
;
alarm for FiO
2
< 18%
Nitrous Oxide (N
2
O)
Measurement range:
0 to 100%
Accuracy:
±3 vol %*
FiN
2
O > 82% alarm
Anesthetic Agent (AA)
Halothane, Isoflurane, Enflurane
Measurement range:
0 to 6%
Accuracy:
±0.2 vol %*
Sevoflurane
Measurement range:
0 to 8%
Accuracy:
±0.2 vol %*
Desflurane
Measurement range:
0 to 20%
Accuracy:
0 to 5% ±0.2 vol %*
5 to 10% ±0.5 vol %
10 to 20% ±1 vol %*
Waveform displayed
MAC value displayed
Identification threshold:
0.15 vol %*
Agent mixture detection
Adjustable high and low alarm limits for EtAA, FiAA
Patient Spirometry
™
(available in Datex-Ohmeda Anesthesia Monitor module bay)
Pressure-volume loop
Flow-volume loop
Airway pressure and flow waveforms
Adjustable low and high alarm limits for P
peak
, PEEP
tot
and
MV
exp
Alarms for MV
exp
<< MV
insp
and for MV
exp
low. Detection
through D-lite
®
or Pedi-lite
®
flow sensor and gas sampler
with following specifications:
D-lite
Pedi-lite
Respiration rate:
4 to 35
4 to 50
breaths/min
breaths/min
Tidal volume
Measurement range:
150 to
15 to
2000 mL
300 mL
Accuracy*:
±6% or 30 mL ±6% or 4 mL
Minute volume
Measurement range:
2 to 20 L/min 0.5 to 5 L/min
Accuracy*:
±6%
±6%
Airway pressure
Measurement range:
-20 to
-20 to
+100 cm H
2
O +100 cm H
2
O
Accuracy*:
±1 cm H
2
O
±1 cm H
2
O
Display units:
cm H
2
O, mmHg, kPa, mbar, hPa
Flow
Measurement range:
1.5 to
0.25 to 25 L/min
100 L/min
I:E
Measurement range:
1:4.5 to 2:1
Compliance
Measurement range:
4 to 100
1 to 100
mL/cm H
2
O
mL/cm H
2
O
Airway resistance
Measurement range:
0 to 40 cm H
2
O/L/s
Sensor specifications
D-lite
Pedi-lite
Dead space:
9.5 mL
2.5 mL
Resistance
at 30 L/min:
0.5 cm H
2
O
at 10 L/min:
1.0 cm H
2
O
Gas exchange**
(available in Datex-Ohmeda Anesthesia Monitor module bay)
VO
2
:
Oxygen consumption
VCO
2
:
Carbon dioxide production
Measurement range:
20 to 1000 mL/min
Respiration rate range:
4 to 35 bpm
Accuracy
FiO
2
< 65%:
±10% or 10 mL
65% < FiO
2
< 85%:
±15% or 15 mL
Detection through D-lite flow sensor or Pedi-lite flow sensor
and gas sampler (see the measurement ranges and sensor
specifications above).
* Typical value
** Measurement not valid with O
2
and N
2
O mixtures
5