O

2

waveform

Measurement range:

0 to 100 %

Accuracy:

±2 vol %*

Datex-Ohmeda differential paramagnetic sensor

Adjustable low and high alarm limits for FiO

2

and EtO

2

;

alarm for FiO

2

< 18%

Nitrous Oxide (N

2

O)

Measurement range:

0 to 100%

Accuracy:

±3 vol %*
FiN

2

O > 82% alarm

Anesthetic Agent (AA)

Halothane, Isoflurane, Enflurane

Measurement range:

0 to 6%

Accuracy:

±0.2 vol %*

Sevoflurane

Measurement range:

0 to 8%

Accuracy:

±0.2 vol %*

Desflurane

Measurement range:

0 to 20%

Accuracy:

0 to 5% ±0.2 vol %*
5 to 10% ±0.5 vol %
10 to 20% ±1 vol %*

Waveform displayed

MAC value displayed

Identification threshold:

0.15 vol %*

Agent mixture detection

Adjustable high and low alarm limits for EtAA, FiAA

Patient Spirometry

(available in Datex-Ohmeda Anesthesia Monitor module bay)

Pressure-volume loop

Flow-volume loop

Airway pressure and flow waveforms

Adjustable low and high alarm limits for P

peak

, PEEP

tot

and

MV

exp

Alarms for MV

exp

<< MV

insp

and for MV

exp

low. Detection

through D-lite

®

or Pedi-lite

®

flow sensor and gas sampler

with following specifications:

D-lite

Pedi-lite

Respiration rate:

4 to 35

4 to 50

breaths/min

breaths/min

Tidal volume

Measurement range:

150 to

15 to

2000 mL

300 mL

Accuracy*:

±6% or 30 mL ±6% or 4 mL

Minute volume

Measurement range:

2 to 20 L/min 0.5 to 5 L/min

Accuracy*:

±6%

±6%

Airway pressure

Measurement range:

-20 to

-20 to

+100 cm H

2

O +100 cm H

2

O

Accuracy*:

±1 cm H

2

O

±1 cm H

2

O

Display units:

cm H

2

O, mmHg, kPa, mbar, hPa

Flow

Measurement range:

1.5 to

0.25 to 25 L/min

100 L/min

I:E

Measurement range:

1:4.5 to 2:1

Compliance

Measurement range:

4 to 100

1 to 100

mL/cm H

2

O

mL/cm H

2

O

Airway resistance

Measurement range:

0 to 40 cm H

2

O/L/s

Sensor specifications

D-lite

Pedi-lite

Dead space:

9.5 mL

2.5 mL

Resistance

at 30 L/min:

0.5 cm H

2

O

at 10 L/min:

1.0 cm H

2

O

Gas exchange**

(available in Datex-Ohmeda Anesthesia Monitor module bay)

VO

2

:

Oxygen consumption

VCO

2

:

Carbon dioxide production

Measurement range:

20 to 1000 mL/min

Respiration rate range:

4 to 35 bpm

Accuracy

FiO

2

< 65%:

±10% or 10 mL

65% < FiO

2

< 85%:

±15% or 15 mL

Detection through D-lite flow sensor or Pedi-lite flow sensor
and gas sampler (see the measurement ranges and sensor
specifications above).

* Typical value
** Measurement not valid with O

2

and N

2

O mixtures