Ventilator components, Anesthetic agent delivery, Compact airway modules – GE Healthcare Avance User Manual

Page 4

background image

Ventilator Components

Flow transducer

Type:

Variable orifice flow sensor

Dimensions:

22 mm OD and 15 mm ID

Location:

Inspiratory outlet and
expiratory inlet

(Optional autoclavable sensor available)

Oxygen sensor

Type:

Optional galvanic fuel cell or
paramagnetic with MGAI option

Ventilator screen

Display size:

31 cm/12.1 inch diagonal

Pixel format:

800 (H) x 600 (V)

Battery backup

Backup power:

Demonstrated battery time
under typical operating
conditions is 90 + minutes
when fully charged.
Battery time under extreme
conditions is 30 minutes.

Battery type:

Internal rechargeable sealed
lead acid

Communication ports

RS-232C compatible serial interface

Ethernet

Datex-Ohmeda device interface solutions port

USB port

Anesthetic Agent Delivery

Delivery

Vaporizers:

Tec

®

5, Tec 6 Plus, Tec 7

Number of positions:

2

Mounting:

Tool-free installation Selectatec

®

manifold interlocks and
isolates vaporizers

Compact Airway Modules

General

M-CAiO, M-CAiOV, M-CAiOVX module software
version 3.2 or higher; E-CAIO, ECAiOV, E-CAiOVX

Size (WxDxH):

75 x 215 x 112 mm/
2.9 x 8.4 x 4.4 in

Weight:

1.6 kg/3.7 lb

Sampling rate:

200 mL/min ±20 mL

Automatic compensation for atmospheric pressure variation
(500 to 800 mmHg) temperature and CO

2

/N

2

O and CO

2

/O

2

collision broadening effect. Parameter display update
interval typically breath-by-breath. Functional alarms for
blocked sample line, D-fend check and D-fend replacement.

Non-disturbing gases:

Ethanol, acetone, methane, nitrogen, nitric oxide, carbon
monoxide, water vapor:

Maximum effect
on readings:

CO

2

< 0.2 vol %; O

2

< 2 vol %

Carbon dioxide (CO

2

)

EtCO

2

:

End-tidal CO

2

concentration

FiCO

2

:

Inspired CO

2

concentration

CO

2

waveform

Measurement range:

0 to 15%
(0 to 15 kPa, 0 to 113 mmHg)

Accuracy:

±0.3 vol %*

Datex-Ohmeda infrared sensor

Adjustable low and high alarm limits for EtCO

2

and FiCO

2

Respiration rate (RR)

Measurement range:

4 to 60 breaths per second

Detection criteria:

1% variation in CO

2

Adjustable low and high alarm limits for respiration rate;
alarm for apnea

Patient Oxygen (O

2

)

FiO

2

:

Inspired O

2

concentration

EtO

2

:

End-tidal O

2

concentration

FiO

2

-EtO

2

:

Inspired-expired difference

* Typical value
** Measurement not valid with O

2

and N

2

O mixtures

4

M1042385-2.qxd:M1042385-1 19.11.2008 10:48 Sivu4