Electrical Specifications

Current leakage

100/120 V:

< 300μA

Power

Power input:

100-120 Vac, 50/60 Hz

Power cord:

Length:

5 m/16.4 ft

Rating:

10A @ 220 Vac or
15A @ 120 Vac

Inlet/outlet modules (100V)

System circuit breakers:

15A

Outlets (optional):

3 outlets on back, 2-2A,
1-4A individual breakers,
isolation transformer standard

Pneumatic Specifications

Auxiliary common gas outlet (optional)

Connector:

ISO 22 mm OD and 15 mm ID

Gas supply

Pipeline input range:

280 kPa to 600 kPa/
41 psig to 88 psig

Pipeline connections:

DISS-male, DISS-female,
DIN 13252, AS4059, BSPP 3/8,
S90-116, or NIST
All fittings available for O

2

,

N

2

O, and Air, and contain

pipeline filter and check valve.

Cylinder input:

Pin indexed in accordance with
CGA-V-1 or DIN (nut and gland);
contains input filter and check valve.
Note: Maximum 2 cylinders

Primary regulator
diaphragm minimum
burst pressure:

2758 kPa/400 psig

Primary regulator
nominal output:

≤ 345 kPa/50 psig
Pin indexed cylinder and
DIN cylinder connections

O

2

controls

Method:

N

2

O shut off with loss of

O

2

pressure

Supply failure alarm:

Range: 193 kPa to 221 kPa/
28 psig to 32 psig
Sounds at maximum volume
every 10 seconds

O

2

flush:

Range: > 35 L/min

Alternate O

2

(safety flow)

Range:

500 mL/min minimum to 10 L/min

Indicator:

Flow tube

Indicator accuracy:

±5% full scale

Fresh gas

Flow range:

0 and 150 mL/min to 15 L/min
(minimal flow capable)

Total flow accuracy:

±10% or ±40 mL/min of setting
(larger of)

O

2

flow accuracy:

±5% or ±20 mL/min of setting
(larger of)

Balance gas
flow accuracy:

±5% or ±20 mL/min of setting
(larger of) Air/N

2

O

O

2

concentration range:

25% to 100%

O

2

concentration

accuracy:

±8% V/V for flows < 400 mL/min
±5% V/V for flows > 400 mL/min
to 1 L/min
±2.5% setting for flows > 1 L/min

Electronic mixer
response time:

500mS (10% to 90% flow step)

Compensation:

Temperature and atmospheric
pressure compensated to
standard conditions of 20°C
and 101.3 kPa

Hypoxic guard:

Electronic

Materials

All materials in contact with patient breathing gases are free
of natural rubber latex.

Environmental Specifications

System operation

Temperature:

10° to 40°C/50° to 104°F

Humidity:

15 to 95% relative humidity
(non-condensing) per IEC 68-2-3

Altitude:

-440 to 3565 m/
500 to 800 mmHg

M1042385-2.qxd:M1042385-1 19.11.2008 10:48 Sivu6