Ventilator accuracy – GE Healthcare Avance User Manual

Page 3

background image

SIMV–PC+PSV (increments of
1 breath per minute)

Inspiratory/
expiratory ratio:

2:1 to 1:8 (increments of 0.5)

Inspiratory time:

0.2 to 5.0 seconds
(increments of 0.1 seconds)
(SIMV and PSV Pro)

Trigger window:

0 to 80% (increments of 5%)

Flow trigger:

1 to 10 L/min
(increments of 0.5 L/min)
0.2 to 1 L/min
(increments of 0.2 L/min)

Inspiration
termination level:

5 to 50% (increments of 5%)

Inspiratory Pause range:

0-60%

Positive End Expiratory Pressure (PEEP)

Type:

Integrated, electronically
controlled

Range:

OFF, 4 to 30 cm H

2

O

(increments of 1 cm H

2

O)

Ventilator performance

Pressure range at inlet:

240 kPa to 700 kPa/
35 psig to 100 psig

Peak gas flow:

120 L/min + fresh gas flow

Flow valve range:

1 to 120 L/min

Flow compensation range: 200 mL/min to 15 L/min

Ventilator Accuracy

Delivery/monitoring accuracy

Volume delivery:

> 210 mL = better than 7%
< 210 mL = better than 15 mL
< 60 mL = better than 10 mL

Pressure delivery:

±10% or ±3 cm H

2

O

PEEP delivery:

±1.5 cm H

2

O

Volume monitoring:

> 210 mL = better than 9%
< 210 mL = better than 18 mL
< 60 mL = better than 10 mL

Pressure monitoring:

±5% or ±2 cm H

2

O

Alarm settings

Tidal volume (V

TE

):

Low: OFF, 0 to 1500 mL
High: 20 to 1600 mL, OFF

Minute volume (V

E

):

Low: OFF, 0 to 10 L/min
High: 0 to 30 L/min, OFF

Inspired oxygen (FiO

2

):

Low: 18 to 100%
High: 19 to 100%, OFF

Apnea alarm:

Mechanical ventilation ON:
< 5 mL breath measured
in 30 seconds

Mechanical ventilation OFF:
< 5 mL breath measured
in 30 seconds

Low airway pressure:

4 cm H

2

O above PEEP

High pressure:

12 to 100 cm H

2

O

(increments of 1 cm H

2

O)

Sustained airway pressure: Mechanical ventilation ON:

P

limit

< 30 cm H

2

O, the

sustained limit is 6 cm H

2

O

P

limit

30 to 60 cm H

2

O,

the sustained limit is 20% of P

limit

P

limit

> 60 cm H

2

O, the

sustained limit is 12 cm H

2

O

PEEP and mechanical
ventilation ON:
Sustained limit increases by
PEEP minus 2 cm H

2

O

Mechanical ventilation OFF:
P

limit

≤ 60 cm H

2

O,

the sustained limit is 50% of P

limit

P

limit

> 60 cm H

2

O, the

sustained limit is 30 cm H

2

O

Subatmospheric pressure: Paw < -10 cm H

2

O

Alarm silence
countdown timer:

120 to 0 seconds