Appendix – American Magnetics 05200PS-430-601 Integrated Power Supply System User Manual
Page 204
186
Rev. 5
Appendix
Model 430 Firmware Upgrade via FTP
8.
Select the Model430.exe file from the Local Site (left side) and drag
it to the open Upgrade folder on the Remote Site (right side)
1
.
9.
Turn off the Model 430.
10. Close the FTP program.
Note
This completes the firmware upgrade. When the Model 430 power is
turned on again after powering down, the firmware will
automatically upgrade based on the Model430.exe file just placed in
its Upgrade folder. View the “Loading. . . . .” screen as the Model
430 is “booting” up: the new Firmware Version will momentarily
2
be displayed.
11. Reboot the Model 430 Programmer by switching the power switch
on the front panel off for 10 seconds and then turning the switch
back on.
12. Verify that after approximately 20 seconds, the Model 430 display
briefly shows the new (upgraded) firmware version
3
:
1. Alternatively, copy
Model430.exe
from the Local Site and paste it to
the open folder on the Remote Site.
2. Display time may vary depending on network speed - refer to
3. Version 1.59 is used only for purposes of this example.
AMI Model 430 Programmer
Firmware Version: 1.61