41

Operation

Shift Key Commands : Voltage Limit

If the Current Limit is raised above the Magnet Current Rating, it

will be ignored and the actual Magnet Current Rating will govern.

3.7.2

Voltage Limit SHIFT-key

Use of the

VOLTAGE LIMIT

SHIFT-key provides a menu for setting the

limit for output voltage for the power supply the Model 430 Programmer

controls. This value should be set to a high enough value so that under

normal conditions, the Voltage Limit is never reached. The value can be

set by using either the numeric keypad per section 3.3 on page 29 or the

fine adjust knob (section 3.4 on page 31). Note that the voltage drop in the

leads must be accounted for when setting the Voltage Limit, as well as the

voltage drop of an energy absorber if one is used (see section 3.12 on

page 79 for details of how to determine the appropriate Voltage Limit).

The Voltage Limit may be set less than or equal to the maximum output

voltage of the power supply.

1

If Voltage Limit becomes active while ramping, it will be indicated by a

reverse illumination character “V” for the status indicator.

Once the Voltage Limit function becomes active, the current, and therefore

field, will no longer be ramping linearly with time as the voltage available

to charge the magnet will be reduced as the total loop voltage will be

limited. As the IR drop of the leads increased with current, the voltage

available to charge the magnet will be reduced.

3.7.3

Reset Quench SHIFT-key

The

RESET QUENCH

SHIFT-key is used whenever a quench detection has

occurred and is being indicated on the display (example shown below).

2. Also if the number of segments is increased, the new segments are added to the

upper end of the ramp range, and default to the ramp rate of the previous segment
with the range of +/-Limit to +/-Limit until set up.

1. Refer to Table 3-2 on page 49.

+50.00 A - Voltage Limit (V)

+0.50 Vs

±2.000

+40.92 A

Mode: Ramping

+2.50 Vs

V

PSwitch: ON

+0.00 A - Quench Detect @ +45.81 A
+0.00 Vs

PSwitch Heater: ON