beautypg.com

Installation – American Magnetics 05100PS-430-601 High Stability Integrated Power Supply System User Manual

Page 39

background image

Rev. 3

21

Installation

Power-Up Procedure

4. When prompted by the Model 430 Programmer, energize the power

supply and press ENTER on the Model 430 Programmer.

Note

Remember to adjust the programmer voltage limit settings as

necessary to account for the additional voltage (5 V) required to

operate the system with a Model 601 installed.

Warning

All power supply parameters, both hardware and software, have

been set at the AMI factory. Power supply control, with the exception

of powering

ON

and

OFF

, is done by way of the AMI Model 430

Power Supply Programmer. No field adjustments or reconfiguration

of the power supply should be attempted in the field unless

specifically described in this document or recommended in writing

by an AMI Technical Support Representative.

5. Enter a stability setting of 100%. Refer to sections 3.3 on page 29

and 3.7.9 on page 43.

6. Verify the various setup menu values for the system (with the

exception of the stability setting, which is to be temporarily left at

100%). If the power supply system was purchased with an AMI

magnet, AMI has preset the setup menu values for proper

operation. See sections 3.3, 3.5, 3.9 and 3.10 for more discussion of

the setup menu values and their entry into the Model 430

Programmer.

7. Set the Model 430 Programmer to display current (rather than

field). Refer to sections 3.2.1 and 3.7.5.

8. Set the ramp rate to 1 A/sec. Refer to sections 3.3 on page 29 and

3.7.1 on page 37.

9. Set the target current to 10 A. Refer to sections 3.3 on page 29 and

3.6.2 on page 35.

10. If a Persistent Switch is installed, set the PSw P/S Ramp Rate to 10

A/sec. Refer to paragraph 3.10.2.11 on page 60

11. Initiate ramping to the target current by pressing the

RAMP /

PAUSE

key (status indicator changes from

P

to ).