Remote-mounted Units
1.) Open the interconnection housing.
2.) Disconnect the probe wiring from the terminal block.
3.) Lower the process fluid level below the bottom of the probe. If this is not possible,
remove the probe from the process vessel. DO NOT remove any coating or
residue present on the probe.
4.) Use an Ohmmeter to measure the resistance between the center of the BNC
Connection at the top of the probe and the probe mounting nut.
5.) If the resistance is below 40Meg
Ω wipe any coating or residue off the probe, and
repeat the measurement. If the reading is still below 40Meg
Ω contact the factory.
See also other documents in the category American Magnetics Equipment:
- 12200PS DC Power Supply (CE-Marked) (58 pages)
- MAxes Multi-Axis Magnet systems for Vector Fields (4 pages)
- 05200PS-430-601 High Stability Integrated Power Supply System (226 pages)
- Current Leads for Cryogenic Systems Brochure (8 pages)
- 150A Portable Liquid Helium Level Meter (CE-Marked) (7 pages)
- AMI Liquid Helium Level Sensors (7 pages)
- 12100PS DC Power Supply (CE-Marked) (53 pages)
- 187 Self-Compensating Liquid Level Controller (75 pages)
- 286 Multi-Sensor Liquid Level Instrument (CE-Marked) (116 pages)
- 601 Energy Absorber (CE-Marked) (7 pages)
- 05100PS-430-601 Integrated Power Supply System (224 pages)
- 4Q06125PS-430 Integrated Power Supply System (216 pages)
- SVTI-CH6022-2 (2 pages)
- OPMX-141-1 (2 pages)
- 05500PS-430-601 High Stability Integrated Power Supply System (228 pages)
- 188CPS Point-Sensing Instrument (29 pages)
- MX-4.5_9-30 (2 pages)
- 4Q05100PS Four-Quadrant Power Supply (7 pages)
- AMI Superconducting Magnets (25 pages)
- Current Leads for Cryogenic Systems Manual (10 pages)
- SD-6-15 (2 pages)
- CRYOGEN FREE SUPERCONDUCTING MAGNET SYSTEMS (4 pages)
- 4Q06250PS-430 Integrated Power Supply System (216 pages)
- 4Q12125PS-430 Integrated Power Supply System (216 pages)
- 430 Power Supply Programmer (4 pages)
- 110 Liquid Helium Level Meter (CE-Marked) (30 pages)
- 420 Power Supply Programmer (134 pages)
- 05100PS-430-601 High Stability Integrated Power Supply System (226 pages)
- 4Q12125PS-430 High Stability Integrated Power Supply System (216 pages)
- 4Q06125PS-430 High Stability Integrated Power Supply System (216 pages)
- 4Q1005PS-430 Integrated Power Supply System (214 pages)
- 160 Liquid Helium Level Superconducting Dipstick (4 pages)
- 05200PS-430-601 Integrated Power Supply System (224 pages)
- 4Q1010PS-430 Integrated Power Supply System (214 pages)
- 05300PS-430-601 High Stability Integrated Power Supply System (226 pages)
- 4Q06250PS-430 High Stability Integrated Power Supply System (216 pages)
- 185 & 186 Liquid Level Instruments (CE-Marked) (82 pages)
- 03300PS-430-601 High Stability Integrated Power Supply System (226 pages)
- 135 & 136 Liquid Helium Level Instruments (CE-Marked) (72 pages)