Power-Up Procedure

9. Set the target current to 10 A. Refer to sections 3.3 on page 31 and

3.6.2 on page 37.

10. If a Persistent Switch is installed, set the PSw P/S Ramp Rate to 10

A/sec. Refer to paragraph 3.10.2.11 on page 62

11. Initiate ramping to the target current by pressing the

RAMP /

PAUSE

key (status indicator changes from

P

to ).

12. The system should ramp to 10 A in approximately 10 seconds.

Verify this is the case (if a PSwitch is installed and in the cooled

state, ramp time to 10 A should be slightly less than 2 seconds).

Note

With an energy absorber unit is connected, the Model 430

Programmer may take significantly longer to ramp the current to 10

A. The Model 430 must first develop a supply output voltage to

overcome the forward voltage drop of a connected energy absorber.

During actual magnet operation, the presence of an energy absorber

will not significantly delay the ramping operation since the Model

430 control gain is increased by orders of magnitude when an

inductive load is connected (unless stability setting is 100%).

13. When the target current is achieved, the

FIELD AT TARGET

LED

will be illuminated. The display should show “

+10.00 A -

indicating that the Model 430 Programmer is in the holding mode

at the target current value (+10.00 A).

Note

There may be a discrepancy between the current shown on the power

supply display

1

and the current displayed on the Model 430

Programmer. The current measurement system incorporated in the

Model 430 is more accurate than the power supply shunt.

14. Verify that the power supply output current displays indicate that

a total of approximately 10 A is being supplied to the load (which is

only the cabling in this case).

15. Set the target current to the Current Limit value. Refer to section

3.10.2.4 on page 58 to determine the Current Limit value. After the

new target current value is entered, the Model 430 Programmer

should ramp automatically to the new setting.

16. When the new target current value is reached, the power supply

current display (if provided) should also indicate the new value.

1. Not all power supplies have a local current readout.