Installation

Magnets w/o Persistent Switch

magnets without a persistent switch, and 2) operation on a short-circuit or

low resistance load.

2.7 Superconducting Magnets with No Persistent Switch

An external stabilizing resistor for superconducting magnets without a

persistent switch is no longer required

1

. However, these systems do

require a specific Model 430 Programmer stability setting based on the

magnet inductance as follows:

For magnet inductance <= 100 Henries (H):

Stability Setting = (100 - H)

For magnet inductance > 100 Henries:

Stability Setting = 0

2.8 Short-Circuit or Resistive Load

If operating with a short-circuit as a load without the presence of a

superconducting magnet, the Model 430 Programmer must be manually

configured for stability. Normally, when the persistent switch heater is

deactivated, the Model 430 Programmer sees essentially a short-circuit

load since the persistent switch shunts all current flow away from any

connected magnet. Therefore, one method of operating a short-circuit is to

indicate that a persistent switch is present, with the persistent switch

heater deactivated.

The preferred method is to indicate that a persistent switch is not present

(see section 3.10.2.6 on page 60) and adjust the stability setting (see

section 3.10.2.1 on page 54) to control the load. A stability setting of 100%

will always allow control of a short-circuit as the load, regardless of the

state of the persistent switch heater.

If the resistance of the load is increased, the stability setting must be

decreased to improve the transient response of the system. If the current

appears to lag, then decrease the stability setting until the system is

responsive. If the current appears to oscillate, increase the stability setting

until the oscillations are damped.

Note

If you have purchased a superconducting magnet with the Model

430 Programmer, AMI will normally provide a recommended

stability setting for optimal operation of the magnet system. If you

1. Effective with Model 430 firmware version 1.62.