Page 9

background image

LED light will blink 3 seconds on and 1 second off and will not stop

blinking until the “data code” is read. The blinking will prompt the

user to contact the technician.

Instruct the customer to read the data code as follows:
To view the “data code”, which is the usage time in hours, push and

hold the MODE button for three seconds. A code will show up on

the display for five seconds and the green light will flash. No other

icons will turn on. After five seconds, the display will return to the

prescribed pressure setting and the green light will stop flashing.

The customer can restart the auto compliance mode by pressing

and holding the mode and up arrow for three seconds. The

actual times will not reset.

To reset the Auto Compliance mode, reset the usage time as

described above.

NOTE: The number of weeks for compliance should be set to

the requirements of the prescribing physician or to the

reimbursement criteria of the payer. Auto compliance is

designed to be an early warning system to the dealer of

issues which may result in less than prescribed usage of

CPAP. This is a first alert of potential clinical problems as well

as potential reimbursement issues.

NOTE: During initial patient evaluations, the unit should

remain plugged into the power source. Unplugging from

the power source will cause the auto compliance feature

to be reset!

VIEW TOTAL TIME:

The TIME icon will appear on the LCD and the total time in

hours will be displayed. The total time for Tier 2 is the time the

device is plugged in. Total time for Tier 3 is since unit was set-up

for a new patient. This value should be zeroed for new patients. To

zero, press and hold up and down arrows simultaneously for

3 seconds until the value reads 0, press mode.

RemRest 9