Activeforever RemRest 901 CPAP Machine User Manual
Page 8

TO VIEW OR RESET THE USAGE TIME:
The USE and TIME icons will flash on the LCD and the
usage time the device was in use by the patient will be displayed on
the LCD. This value should be zeroed for new patients to remove
factory test hours.
Press the up and down arrows at the same time to reset the
usage time. Once the desired setting is reached, press the mode
button to save the setting.
TO SET THE RAMP STARTING PRESSURE:
The default setting for the starting ramp pressure is 3.5 cmH2O.
The
LO icon will flash and the starting ramp pressure will be
displayed on the LCD. Press the up arrow or down arrow
to adjust the pressure to the desired start ramp pressure. The
pressure will adjust in increments of .1cmH2O. To move through the
settings more quickly, hold down the up or down arrow
button. Once the desired setting is reached, press the mode button
to save the setting.
TO SET THE RAMP TIME:
The
RAMP and
TIME icons will flash on the LCD and the
time in minutes will be displayed on the LCD. Press the up arrow
or down arrow to adjust the ramp time. The time will adjust in
5 minute increments from 5 to 50 minutes. Once the desired setting
is reached, press the mode button to save the setting.
TO SET THE AUTO COMPLIANCE:
The code “AC0” is displayed on the LCD indicating the factory
default setting of zero weeks. Press the up arrow or down
arrow to adjust the number of weeks from 1 to 26. Adjusting the
display to read AC0 or zero weeks turns off the auto compliance
feature. Once the desired setting is reached, press the mode button
to save the setting. The unit will perform a weekly check to confirm
28 hours of usage time per week (4 hours/day) for one to 26
weeks. If the client has failed to use the device, then the green
8
RemRest