Activeforever RemRest 901 CPAP Machine User Manual
Page 7

RemRest 7
To enter the Technician Mode, plug one end of the AC power cord
into the AC inlet plug located on the back of the RemRest device.
Plug the other end into the wall socket. The light will turn on for two
seconds then will turn off. The LCD will also turn on completely
(as shown above) for 2 seconds and will then display prescription
pressure.
With the motor blower off,
• hold down the “up” arrow and the “down” arrow
• simultaneously push and hold on/off button for two seconds
• the light should start blinking 1/2 second on and 1/2 second off
NOTE: First time device is put into service with a new patient, run
time, usage time and total time must be reset to insure accurate
compliance information.
NOTE: To advance through the different modes, press mode button
repeatedly until in the desired mode.
TO SET THE PRESCRIPTION PRESSURE:
The default setting is 10cmH2O. Press the up arrow or down
arrow to adjust the pressure to the prescribed setting. The
HI
will flash on the LCD and the pressure will be displayed cmH2O as
xx.x. The pressure will adjust in increments of 0.1cmH2O. To move
through the settings more quickly, hold down the up or down
arrow button. Once the desired setting is reached, press the mode
button to save the setting and enter next mode.
TO VIEW OR RESET THE RUN TIME:
The RUN and TIME icons will flash on the LCD and the
run time will be displayed on the LCD. Run time is the number of
hours the device is powered on. This value should be zeroed for new
patients to remove factory test hours.
Press the up and down arrows at the same time to reset the
run time to 0. Once the desired setting is reached, press the mode
button to save the setting.